Related Experiment Video
Updated: Aug 27, 2026

Mimicking and Measuring Occlusal Erosive Tooth Wear with the "Rub&Roll" and Non-contact Profilometry
Published on: February 2, 2018
In vivo assessment of the Optipen optical reflectometer for detecting enamel surface changes following an erosive
Ana Luiza Bogaz Debortolli1, Franciny Querobim Ionta1, Ana Laura Marques Xavier1
1Department of Pediatric Dentistry, Orthodontics and Public Health, Bauru School of Dentistry, University of São Paulo, Bauru, SP, Brazil.
Objectives:
To evaluate whether the Optipen optical reflectometer can detect enamel surface changes associated with erosion under clinical conditions.
Materials And Methods:
In this short-term parallel randomized clinical trial, 24 healthy participants were allocated to two groups: Test (T-rinse with a cola beverage for 30 s) and Control (C-rinse with water). Surface reflectance was measured using the Optipen optical reflectometer on the maxillary central incisors at baseline, immediately after rinsing, and after 2 h of salivary exposure, and expressed as the percentage of surface reflection intensity.
Results:
The test group showed significantly lower reflectance values immediately after the erosive challenge (65.4) compared with the control group (102) (p < 0.05). After 2 h of salivary exposure, no significant differences were observed between groups (T = 85.7 and C = 97.8).
Conclusions:
The Optipen optical reflectometer detected immediate changes in enamel surface reflectance in vivo following a single erosive challenge.
Clinical Relevance:
This device may be useful for detecting immediate enamel surface changes following erosion. Further studies are needed to optimize reflectance measurements after salivary exposure.