Disparities in Blood Pressure Measurement
Ayan Ali1, Liann Abu Salman2, Jordana B Cohen3,4
1Cardiology Division, Department of Medicine, Feinberg School of Medicine, Northwestern University, Chicago, IL, USA.
Purpose Of Review:
We aimed to examine how inaccuracies and inequities in blood pressure (BP) measurement across office, ambulatory, home, and emerging cuffless modalities contribute to disparities in hypertension diagnosis and control, and to identify strategies to improve measurement equity.
Recent Findings:
Disparities arise from nonstandardized office technique, inappropriate cuff sizing, limited access to validated home and ambulatory devices, reimbursement and implementation barriers, and underrecognition of masked and nocturnal hypertension in populations at higher risk. Cuffless technologies may expand access and reduce cuff-related barriers, but current devices have important limitations in validation, calibration, real-world accuracy, and performance across diverse populations. Improving equity requires standardized measurement, access to appropriately sized validated devices and out-of-office monitoring, inclusive device validation, and reimbursement and implementation strategies that reach underrepresented and under-resourced communities. Future research should test whether these approaches reduce diagnostic misclassification and improve BP control and cardiovascular outcomes across populations.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...