Validation study on cuffless ambulatory blood pressure monitoring method using bio-impedance, impedance cardiography
Myrte Schoenmakers1,2,3, Martin Gevonden1,2,3, Bruno Sauce1,4
1Department of Biological Psychology, Vrije Universiteit Amsterdam, Van der Boechorststraat 7, 1081 BT, Amsterdam, The Netherlands.
Purpose:
Continuous, non-invasive blood pressure (BP) monitoring is essential for understanding cardiovascular dynamics in real-world settings. This study evaluated the reliability and validity of a cuffless method for ambulatory BP monitoring using pulse wave travel times derived from bio-impedance, impedance cardiography, and electrocardiography. We examined whether this method can accurately estimate systolic and diastolic BP cross varying conditions.
Methods:
Forty-one young, healthy participants completed a protocol involving postural changes, physical activity, and mental stress tasks. Pulse arrival time (PAT) and pulse transit time (PTT) were extracted, and person-specific models estimated BP using (1) all data of the person ('double-dipping') - used in most previous studies -, (2) half of a person's data (calibration), or (3) person-invariant parameters (calibration-free) which would enhance practical utility. Cuff-based BP served as the reference. Test-retest reliability, and construct and convergent validity were assessed.
Results:
Test-retest reliability was high for PAT- and PTT-based BP for all approaches (ICC > 0.7), comparable to cuff-based measurements. Construct validity was generally poor: BP changes during physical activity were detected, but responses to mental stressors were not. Only double-dipping showed acceptable convergent validity, likely reflecting overfitting. In the calibration approach, convergence with cuff-based BP was moderate (MAE = 10.1-12.9 mmHg), with 59.2-68.5% of estimates within 10 mmHg of reference values. The strongest deterioration of convergentvalidity occurred with person-invariant models.
Conclusion:
Although cuffless pulse wave travel time-based BP estimators showed strong reliability, validity was insufficient for clinical or stress research use, even with person-specific calibration. Future work should explore extended calibrationperiods before testing in naturalistic settings.
More Related Videos
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure