Herriott-type cavity-enhanced Raman spectroscopy for simultaneous analysis of ten natural gas components
Related Experiment Video
Updated: Oct 30, 2026

Infrared Degenerate Four-wave Mixing with Upconversion Detection for Quantitative Gas Sensing
Published on: March 22, 2019
Xu Dong1, Yue Wang2, Yanli Qin2
1College of Science, Shenyang Ligong University, Shenyang, Liaoning, 110159, PR China; Zolix Instruments Co., Ltd, Beijing, 101102, China.
Abstract:
Real-time monitoring of natural gas composition is essential for reservoir quality evaluation, combustion optimization, and safe transportation. Raman spectroscopy can detect multiple gas species simultaneously through molecular fingerprint signatures, yet weak scattering cross-sections limit its sensitivity in gas-phase applications. Here we present a compact Herriott-type cavity-enhanced Raman spectroscopy (CERS) system in which a 532 nm laser undergoes 128 passes within a 382 mm cavity, yielding a geometric beam path of 49.02 m and a calculated power-weighted enhancement factor of 94.7. The system achieves ppm-level limits of detection for all ten target species in natural gas: 2.8 ppm for H₂, 6.0 ppm for CO, 5.6 ppm for CO₂, 48.8 ppm for CH₄, 6.2 ppm for C₂H₆, 4.7 ppm for C₃H₈, 23.2 ppm for nC₄H₁₀, 3.2 ppm for iC₄H₁₀, 16.0 ppm for nC₅H₁₂, and 7.3 ppm for iC₅H₁₂. DFT calculations at the B3LYP/6-311++G(d,p) level using second-order vibrational perturbation theory (VPT2) anharmonic corrections support the vibrational mode assignments, yielding a mean absolute deviation of 12.5 cm-1. Quantitative analysis of a multi-component mixture yields quantification deviations below 10% for the ten species with RSD below 5%. These results demonstrate the potential of this system for field-deployable natural gas monitoring.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Gas Chromatography: Types of Detectors-II
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...