Technical performance evaluation of mobile x-ray systems for chest imaging
Related Experiment Video
Updated: Nov 12, 2026

Clinical Imaging of Microwave Mammography
Published on: November 14, 2025
N W Marshall1, J Binst2, H Verhoeven2
1Competentiecentrum Medische Stralingsfysica, Radiologie, UZ Leuven, Herestraat 49, 3000 Leuven, Belgium; Medical Physics and Quality Assessment, Katholieke Universiteit Leuven 3000 Leuven, Belgium.
Purpose:
To present both quantitative and test object-based imaging performance measurements for the latest mobile digital radiography (DR) x-ray systems for bedside imaging.
Methods:
Seven mobile DR systems with CsI-based detectors were assessed. The modulation transfer function (MTF) and detective quantum efficiency (DQE) were measured using the RQA8 beam quality at three detector incident air kerma levels (KB). A weighted polynomial model applied to the variance was used to estimate electronic, quantum and structure noise fractions versus KB. Technical image quality was assessed by imaging the Leeds TO12 contrast-detail phantom with 10 cm, 15 cm and 20 cm PMMA, using the x-ray technique factors specified by the applications specialists for the adult Chest AP examination.
Results:
Electronic noise dominated below a mean value of 0.14 µGy, while quantum noise comprised 91.6% ,on average, of total image noise at 2.5 µGy. The spatial frequency for an MTF of 0.1 ranged from 3.2 mm-1 to 4.7 mm-1, with a mean of 3.8 mm-1. DQE at 0.5 mm-1, measured at the RQA8 quality and a target KB value of 2.5 µGy, ranged from 0.46 to 0.64 with a mean of 0.55. Mean nominal threshold contrast for the 2 mm diameter disc (±1 standard deviation) was 1.72% ± 0.23%, 3.66% ± 0.68% and 8.05% ± 2.78%, for 10 cm, 15 cm and 20 cm PMMA, respectively.
Conclusion:
Despite some variation between detectors, overall imaging performance for the systems was good. These data could act as reference values for medical physicists assessing x-ray devices for bedside imaging.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT