Relationship between amylase concentration, L/S ratio and lecithin concentrations in amniotic fluid
Gynecologic and Obstetric Investigation
|January 1, 1979
Summary
The amylase test in amniotic fluid effectively predicts fetal maturity. Lower amylase levels indicate preterm status, while higher levels suggest fetal maturity, but intermediate levels require further testing.
Area of Science:
- Obstetrics and Gynecology
- Clinical Chemistry
- Neonatal Medicine
Background:
- Assessing fetal lung maturity is crucial for optimizing delivery timing.
- Traditional methods like the lecithin/sphingomyelin (L/S) ratio and lecithin concentration can be complex and time-consuming.
Purpose of the Study:
- To evaluate the utility of amniotic fluid amylase concentration as a predictor of fetal maturity.
- To compare amylase levels with established fetal maturity markers such as L/S ratio and lecithin concentration.
Main Methods:
- Amniotic fluid samples (n=110) from 106 uneventful pregnancies were analyzed.
- Amylase concentration, L/S ratio, and lecithin concentration were measured in each sample.
- Statistical analysis was performed to correlate amylase levels with gestational age and phospholipid concentrations.
Main Results:
- Amylase concentrations < 200 U/l significantly correlated with gestational age < 37 weeks, L/S ratio < 2, and lecithin < 3.5 mg% (p<0.001).
- Amylase concentrations > 300 U/l significantly correlated with gestational age > 37 weeks, L/S ratio > 2, and lecithin > 3.5 mg% (p<0.001).
- Amylase levels between 200-300 U/l showed no significant correlation with gestational age or phospholipid levels.
Conclusions:
- Amniotic fluid amylase concentration is a simple, rapid, and cost-effective screening method for predicting fetal maturity.
- Amylase levels < 200 U/l suggest prematurity, while levels > 300 U/l indicate maturity.
- Intermediate amylase values (200-300 U/l) necessitate confirmatory testing using specific fetal maturity assessments.