A radioimmunoassay for serum ferritin
Clinical Chemistry
|January 1, 1977
Summary
A new radioimmunoassay for serum ferritin was developed, offering a sensitive and precise method for detecting iron deficiency. Low ferritin levels (<18 µg/L) reliably indicate iron deficiency anemia.
Area of Science:
- Clinical Chemistry
- Immunology
- Hematology
Background:
- Serum ferritin is a key indicator of iron status.
- Accurate measurement of serum ferritin is crucial for diagnosing iron deficiency.
Purpose of the Study:
- To develop and evaluate a novel radioimmunoassay for serum ferritin.
- To establish reference intervals and assess the assay's diagnostic utility for iron deficiency.
Main Methods:
- Development of a radioimmunoassay using 125I-labeled ferritin and specific antibodies.
- Utilized goat anti-rabbit gamma-globulin and polyethylene glycol for separation.
- Assayed serum samples with a working range up to 500 µg/L and 75 µL sample volume.
Main Results:
- The assay demonstrated high sensitivity (1.5 µg/L) and long-term precision (CV 13%).
- Established reference intervals: men (18-330 µg/L), women >50 years (18-200 µg/L).
- Serum ferritin <18 µg/L correlated with absent bone marrow iron, indicating iron deficiency.
Conclusions:
- The developed radioimmunoassay is a sensitive and precise tool for measuring serum ferritin.
- Serum ferritin levels below 18 µg/L are a reliable indicator of iron deficiency.
- The assay aids in the diagnosis of iron deficiency across different demographic groups.