Low temperature steam and formaldehyde sterilization.
The Journal of Hospital Infection
|September 1, 1983
Summary
Standard steam/formaldehyde autoclaves failed to sterilize Bacillus stearothermophilus spores. Modifications and an improved cycle ensured reliable sterilization, confirmed in a prototype unit.
Area of Science:
- Microbiology
- Sterilization Technology
- Medical Device Engineering
Background:
- Standard low temperature steam/formaldehyde autoclaves are used for sterilization.
- Initial testing revealed failures in eliminating Bacillus stearothermophilus spores, a common challenge organism.
Purpose of the Study:
- To identify reasons for sterilization failures in standard autoclaves.
- To develop and validate modifications for improved autoclave performance.
- To establish reliable sterilization conditions.
Main Methods:
- Testing a standard low temperature steam/formaldehyde autoclave with Bacillus stearothermophilus spore-containing test pieces.
- Analyzing failure modes and implementing autoclave modifications.
- Developing and testing an improved operating cycle.
- Validating performance with a commercially produced prototype autoclave.
Main Results:
- Initial autoclave operation showed failures in killing Bacillus stearothermophilus.
- Modifications and an improved cycle led to greatly improved performance.
- Reliable and consistent sterilization of all test pieces was achieved.
- A similarly modified prototype autoclave also demonstrated effective sterilization.
Conclusions:
- Standard autoclaves require modification for effective sterilization against resistant organisms.
- An improved operating cycle and specific modifications ensure reliable autoclave sterilization.
- The validated modifications are applicable to both standard and prototype autoclaves.