Related Experiment Video
Updated: Jul 25, 2026

07:31
Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Operating characteristics of the Dynaray 18 linear accelerator
Acta Radiologica. Oncology
|January 1, 1983
Abstract:
The operating system of a Dynaray 18 linear accelerator is described. Central axis depth dose data for the 10 MV roentgen ray beam are given and compared with published data. The alignment of the roentgen ray beam and the modifications required to obtain satisfactory uniformity and reliable dosimetry are described. The characteristics of the electron beams are also described, central axis depth dose data are given and the derived estimates of energy are discussed.