Evaluation of the Quantum II and Rapid E identification systems
Journal of Clinical Microbiology
|September 1, 1984
Summary
The API 20E, Rapid E, and Quantum II systems accurately identified Enterobacteriaceae. Quantum II showed higher accuracy, with reproducibility improved after modification, offering cost-effective bacterial identification.
Area of Science:
- Clinical Microbiology
- Bacterial Identification
- Diagnostic Systems Evaluation
Background:
- Accurate identification of Enterobacteriaceae is crucial for patient care and infection control.
- Several automated and semi-automated systems exist for bacterial identification, each with varying performance characteristics.
Purpose of the Study:
- To evaluate the accuracy and reproducibility of the API 20E, Rapid E, and Quantum II identification systems for Enterobacteriaceae.
- To compare the performance of these systems against conventional biochemical tests.
Main Methods:
- 492 clinical Enterobacteriaceae isolates were tested using API 20E, Rapid E, and Quantum II systems.
- Discrepant results were resolved using repeat testing or conventional biochemical methods.
- Reproducibility was assessed by triplicate testing of 40 organisms on Rapid E and Quantum II systems.
Main Results:
- API 20E and Rapid E achieved 94.1% correct identification; Quantum II achieved 97.0% correct identification for Enterobacteriaceae.
- Quantum II correctly identified 83.3% of 48 non-Enterobacteriaceae isolates.
- Aberrant biochemical reactions were the primary cause of misidentifications in Rapid E and Quantum II; Quantum II reproducibility improved post-modification.
Conclusions:
- The API 20E, Rapid E, and Quantum II systems demonstrate good accuracy for identifying Enterobacteriaceae.
- Quantum II exhibited the highest accuracy, with improved reproducibility after photometer modification.
- Rapid E and Quantum II are cost-effective, user-friendly alternatives for bacterial identification.
Related Concept Videos
Mass Analyzers: Common Types
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
Raman Spectroscopy Instrumentation: Overview
A conventional Raman spectrophotometer includes a laser source, a sample holding system, a wavelength selector, and a detector.
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Rapid Identification of Pathogens
MALDI-TOF MS has transformed clinical microbiology by offering a rapid and reliable method for pathogen identification. The traditional approach to microbial identification typically involves time-consuming culture techniques and biochemical tests, which can delay the initiation of appropriate antimicrobial therapy. MALDI-TOF MS avoids these delays by using characteristic ribosomal protein mass patterns of microbial cells, enabling accurate species-level identification within minutes.Principle...
Automated Microbial Diagnostics
Automated diagnostic analyzers have transformed clinical microbiology by providing rapid and reliable methods for pathogen identification and antibiotic susceptibility testing. Among these systems, the Vitek 2 is widely used because it automates the traditionally labor-intensive processes of microbial identification (ID) and antibiotic susceptibility testing (AST), delivering standardized and timely results that are essential for effective patient care.Microbial Identification with ID CardsThe...