Related Experiment Videos
High-performance hydrophobic interaction chromatography of proteins
Abstract:
A new, weakly hydrophobic, high-performance liquid chromatography column has been developed for the separation of native proteins based on their relative hydrophobicities. Starting with a covalently bound, hydrophilic polyamine matrix, packing materials were synthesized through acylation with anhydrides and acid chlorides of increasing chain length to obtain increasingly hydrophobic surfaces. Proteins in aqueous buffers were induced to bind hydrophobically to the columns by the use of high salt concentrations in the mobile phase. Elution was achieved by decreasing the ionic strength of the solvent in a linear gradient. A mixture of cytochrome c, conalbumin, and beta-glucosidase was used as a standard to test the resolving power of newly synthesized columns. On a 4-cm butyrate column, baseline resolution was achieved in 20 min with a gradient of 3.0 mu sodium sulfate in 0.1 M potassium phosphate buffer, pH 7.0, to water. The static loading capacity for each column was determined using a hemoglobin binding assay. Capacities normally ranged between 150 and 180 mg of hemoglobin per gram of support. Since proteins are not denatured in hydrophobic interaction chromatography, enzymes eluted from the column retained enzymatic activity. Samples of alpha-amylase and beta-glucosidase ranging in size from 10 to 200 micrograms were recovered from the butyrate column with greater than 92% enzymatic activity in all cases. In a single trial, the enzyme citrate synthase was recovered from the benzoate column with 92% retention of enzymatic activity.