Related Experiment Videos

Late outcome of patients with Braunwald-Cutter mitral valve replacement

Insights

The Braunwald-Cutter valve shows low embolic complication rates in mitral valve replacement (MVR) patients. Poppet wear is minimal in mitral/tricuspid positions, but poppet escape occurred in the aortic position.

Area of Science:

  • Cardiovascular Surgery
  • Biomaterials Science

Background:

  • The Braunwald-Cutter valve was used for mitral valve replacement (MVR) between 1972 and 1975.
  • Long-term outcomes and complications of this prosthesis were not fully established.

Purpose of the Study:

  • To evaluate the long-term efficacy and safety of the Braunwald-Cutter valve in patients undergoing mitral valve replacement.
  • To assess survival rates, embolic complications, and structural integrity of the prosthesis over a 10-year follow-up period.

Main Methods:

  • Retrospective analysis of 80 patients who received Braunwald-Cutter prostheses for mitral valve replacement.
  • Follow-up duration ranged from 72 to 120 months (mean 84.6 months).
  • Actuarial survival, embolic complication rates, and valve structural integrity were analyzed.

Main Results:

  • Ten-year actuarial survival was 73% for MVR alone and 30% for multiple valve replacement.
  • Linearized embolic rates were 3.2%/year for MVR and 1.5%/year for multiple valve replacement.
  • No significant poppet wear or escape observed in mitral/tricuspid positions after 10 years; poppet escape occurred in the aortic position up to 101 months.

Conclusions:

  • The Braunwald-Cutter valve demonstrates favorable embolic complication rates, comparable to other mechanical prostheses.
  • Elective replacement due to poppet wear/escape is unnecessary for mitral/tricuspid positions.
  • High overall morbidity highlights the risks associated with valve replacement surgery.