Related Experiment Video
Updated: Aug 7, 2026

Multifocal Electroretinograms
Published on: December 4, 2011
Macular electroretinograms: their accuracy, specificity and implementation for clinical use
Abstract:
The pattern and the focal electroretinogram (ERG) are both non-invasive, electrophysiological responses recorded from circumscribed retinal areas and are most easily recorded from the macula. This paper describes how our department has incorporated these tests into our clinical protocol, shows how the recording technique and the method of electrode construction may be improved, and describes the normal limits of the macular responses we obtain. The ERG signal-noise ratio we obtained was better than that of the binocular visual evoked potentials (VEPs) recorded simultaneously. Pattern and focal ERGs, using improved methods of recording, show promise of being a valuable addition to the clinical investigation of subtle maculopathies and some forms of optic nerve dysfunction. Three illustrative cases are described. The first demonstrates normal macular ERG responses with abnormal Ganzfeld ERGs due to peripheral retinal damage. The second reveals differential pattern ERG reduction with normal focal ERG in recent optic neuritis. The third case demonstrates reversible simultaneous loss of Ganzfeld ERGs and macular ERGs in vitamin A deficiency.
More Related Videos
06:34Using the Electroretinogram to Assess Function in the Rodent Retina and the Protective Effects of Remote Limb Ischemic Preconditioning
Published on: June 9, 2015
08:38Electroretinogram Recording for Infants and Children under Anesthesia to Achieve Optimal Dark Adaptation and International Standards
Published on: September 3, 2020