Tissue reaction to a failed Bion hip prosthesis in a dog
Biomaterials, Medical Devices, and Artificial Organs
|January 1, 1984
Summary
A dog experienced a severe inflammatory tissue reaction after a total hip replacement using a Bion elastomer component failed. The implant fracture led to granulomatous inflammation and macrophages filled with Bion material.
Area of Science:
- Biomaterials science
- Veterinary orthopedics
- Tissue engineering
Background:
- Total hip replacement (THR) systems are used to treat hip joint diseases.
- Implant failure can lead to adverse tissue reactions and complications.
- Elastomer components in THR devices require evaluation for biocompatibility and durability.
Purpose of the Study:
- To evaluate the tissue reaction in a canine model to a failed constrained total hip replacement system.
- To investigate the consequences of Bion elastomer component fracture within a THR device.
Main Methods:
- A canine subject received a constrained total hip replacement with Ti-6A1-4V alloy components and a Bion elastomer.
- The implant was explanted at necropsy following device failure.
- Histopathological analysis was performed on the surrounding tissues to assess the inflammatory response.
Main Results:
- The Bion elastomer component of the total hip replacement system was found to have fractured.
- Tissue surrounding the fractured component exhibited granulomatous inflammation with blackened regions.
- Histology revealed a severe chronic inflammatory response, characterized by giant cells and Bion-laden macrophages.
Conclusions:
- Fracture of the Bion elastomer component in a total hip replacement system can elicit a significant adverse tissue reaction.
- The observed granulomatous inflammation and presence of Bion material within macrophages indicate a foreign body response.
- This case highlights potential biocompatibility and mechanical failure issues with elastomer components in orthopedic implants.