Related Experiment Video
Updated: Aug 3, 2026

The Use of the Ex Vivo Chandler Loop Apparatus to Assess the Biocompatibility of Modified Polymeric Blood Conduits
Published on: August 20, 2014
In vivo validation of a cell culture test for biocompatibility testing of urinary catheters
Abstract:
Biocompatibility tests have been compared for their suitability as routine safety tests for urinary catheters. Latex catheters from five manufacturers were tested by each of the following four methods: (1) a cell culture cytotoxicity assay of catheter extracts, (2) intracutaneous injection of the extracts into rabbits, (3) intramuscular implant of catheter pieces into rabbits, (4) catheterization of sheep (mucous membrane irritation). The rabbit intracutaneous and intramuscular tests are both current pharmacopoeial methods for ascertaining the suitability of polymers for medical use. The four tests each showed a cell or tissue response ranging from no detectable change to severe damage, according to the catheter batch or brand, and they each identified the same samples as most toxic and least toxic. However, they differed in sensitivity. The sheep test and the cell culture assay discriminated between catheters of intermediate toxicity and ranked as toxic catheters not identified as such by the two pharmacopoeial tests. The sheep test most closely approximates clinical usage, but is impractical for routine use. The cell culture assay is a suitable alternative. It also has the advantages of a clearly defined endpoint, good sensitivity, reproducibility, speed, and reduced animal usage.

