Strut fractures in De Bakey aortic valves
Scandinavian Journal of Thoracic and Cardiovascular Surgery
|January 1, 1983
Summary
De Bakey aortic valve prostheses implanted in 1973 experienced strut fractures, leading to fatal embolization in one patient. Early recognition and valve replacement are recommended for this specific aortic valve prosthesis type.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Prosthetic Heart Valves
Background:
- The De Bakey aortic valve prosthesis, implanted in 1973, is a mechanical heart valve designed for aortic valve replacement.
- Long-term durability and potential failure modes of early prosthetic heart valves are critical considerations in cardiovascular surgery.
Observation:
- Strut fractures were observed in two out of six patients (33%) who received the De Bakey aortic valve prosthesis.
- Fractures occurred at 7.5 and 9 years post-implantation, indicating a potential long-term mechanical failure.
- One patient experienced fatal cage and ball embolization due to prosthesis failure.
Findings:
- Mechanical failure, specifically strut fracture, is a significant risk associated with the early De Bakey aortic valve prosthesis.
- The time to failure suggests a material fatigue or design limitation over prolonged use.
- Prompt diagnosis and intervention can prevent catastrophic outcomes, as demonstrated by the successful valve replacement in the second patient.
Implications:
- Valve replacement is strongly recommended for patients with the De Bakey aortic valve prosthesis of this specific design and implantation era.
- This case highlights the importance of long-term surveillance for prosthetic heart valves and the need for updated designs.
- Further research into the failure mechanisms of older prosthetic devices is warranted to improve future implantable cardiac technologies.