Skinfold measurements in children. A comparison of Lange and McGaw calipers
Insights
McGaw and Ross plastic calipers provide comparable skinfold measurements to Lange calipers in pediatric patients. These plastic calipers offer an adequate, cost-effective alternative for pediatric body composition analysis.
Area of Science:
- Pediatric Endocrinology
- Human Physiology
- Anthropometry
Background:
- Accurate anthropometric measurements are crucial for assessing pediatric growth and nutritional status.
- Previous studies compared skinfold calipers in adults, but pediatric data is limited, especially for smaller skinfolds common in children.
- Understanding inter-caliper and inter-observer variability in pediatric skinfold assessment is essential for reliable body composition analysis.
Purpose of the Study:
- To compare skinfold measurements obtained using McGaw and Lange calipers in a pediatric population.
- To evaluate inter-observer reliability for both caliper types in pediatric patients.
- To determine if McGaw and Ross plastic calipers are a suitable alternative to Lange calipers for pediatric use.
Main Methods:
- Biceps and triceps skinfold thickness were measured in 240 pediatric patients (age 2 weeks to 18 years).
- Measurements were taken using both McGaw and Lange calipers on the same individuals.
- Two independent investigators performed measurements to assess inter-observer differences.
Main Results:
- McGaw caliper measurements were, on average, 12.6% lower than those from Lange calipers.
- Inter-observer differences were 20.4% for Lange calipers and 17.6% for McGaw calipers (root mean square differences).
- Both caliper types demonstrated variability, but McGaw calipers showed slightly better inter-observer consistency.
Conclusions:
- McGaw and Ross plastic calipers are an adequate substitute for Lange calipers in pediatric settings, particularly when cost is a factor.
- The findings support the use of McGaw calipers for reliable pediatric anthropometry, despite lower absolute values compared to Lange calipers.
- Further research may explore the clinical implications of these measurement differences in pediatric health assessments.
Abstract:
Two types of calipers to measure skinfolds have been compared in adults, but the same type of comparison has not been done in a pediatric population who generally would have smaller skinfolds. Biceps and triceps skinfold measurements were made in 240 pediatric patients ranging in age from 2 weeks to 18 years. Skinfold measurements obtained with a McGaw caliper were on the average 12.6 percent lower than those obtained with a Lange caliper on the same patient. Differences between measurements by two investigators using the same caliper on the same patient averaged 20.4 percent for the Lange caliper and 17.6 percent for the McGaw caliper (root mean square differences). We conclude that McGaw and Ross plastic calipers would be an adequate substitute where cost prohibits the use of Lange calipers.