Related Experiment Video
Updated: Aug 11, 2026

Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry (UPLC-HRMS)
Published on: May 20, 2013
Optimized liquid-chromatographic determination of metronidazole and its metabolites in plasma
Abstract:
Metronidazole and its known metabolites in plasma can be rapidly separated by a "high-pressure" liquid-chromatographic method that can also be adapted for rapid determination of tinidazole. Samples deproteinized with trichloroacetic acid (50 g/L final concentration) undergo isocratic separation on a reversed-phase C18 column eluted with an 8/92 (by vol) mixture of acetonitrile/KH2PO4 (5 mmol/L, pH 3.0). The method is sensitive, reliably detecting as little as 25 micrograms of metronidazole and (or) its metabolites per milliliter of plasma. The detector response varied linearly with concentration for all compounds tested over a wide range (25-500 micrograms/L). Within-day and between-day variation was generally less than 2.5% for all concentrations of all compounds tested. Various other antibiotics tested did not interfere.
More Related Videos
11:25Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021