[Sonographic artifacts of round reflectors]
Summary
This study demonstrates the proximal curved edge artifact in linear-array ultrasound systems using real cases and waterbath simulations. A physical explanation for this common imaging artifact is provided.
Area of Science:
- Ultrasound imaging
- Medical physics
Context:
- Linear-array ultrasound systems are widely used in medical diagnostics.
- The proximal curved edge artifact can affect image quality and diagnostic accuracy.
Purpose:
- To demonstrate the appearance of the proximal curved edge artifact.
- To simulate the artifact using a waterbath.
- To provide a physical explanation for the artifact.
Summary:
- The proximal curved edge artifact was observed in five clinical cases using linear-array systems.
- Simulations in a waterbath successfully replicated the artifact.
- The underlying physical principles causing this artifact are discussed.
Impact:
- Improved understanding of ultrasound artifact generation.
- Potential for enhanced image interpretation and artifact reduction strategies.
- Aids in quality assurance for ultrasound equipment and techniques.