Comparison of noninvasive and direct measurements of intracranial pressure
Abstract:
After creating "fontanelles" in infant rhesus monkeys, we measured intracranial pressure of these monkeys directly with an indwelling cannula and indirectly with a Ladd 1700 sensor. Values obtained with the Ladd sensor placed over the shaven fontanelle closely approximate directly measured pressure in acute experiments in these animals and in human cadavers. Application of the device is simple and application pressure need not always be the same for multiple determinations. The response time is 1.6 cm H2O/sec, which is adequate to record changes in intracranial pressure caused by changes in the partial pressure of CO2 in arterial blood.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...