CAP blood gas survey--first year's experience.
American Journal of Clinical Pathology
|October 1, 1980
Summary
Clinical chemistry laboratories evaluated blood gas analysis precision. While pH showed good interlaboratory precision, PO2 results were less precise, with differences linked to instruments rather than manufacturers.
Area of Science:
- Clinical Chemistry
- Pathology
- Medical Laboratory Science
Background:
- Clinical laboratories perform vital diagnostic tests, including blood gas analysis.
- Interlaboratory variability in blood gas measurements can impact patient care.
- Standardized surveys are crucial for assessing analytical performance.
Purpose of the Study:
- To assess the interlaboratory precision and accuracy of blood gas analysis (pH, PO2, PCO2).
- To identify sources of variability in blood gas measurements among participating laboratories.
- To evaluate the performance of different analytical instruments in a survey setting.
Main Methods:
- Over 500 clinical chemistry/pathology laboratories participated in the survey.
- Participants analyzed provided survey specimens containing pH, PO2, and PCO2.
- Interlaboratory precision was assessed, and interinstrument differences were analyzed using all-methods results.
Main Results:
- pH demonstrated the best interlaboratory precision, while PO2 exhibited the lowest.
- Accuracy evaluation was limited by the absence of an accepted reference method.
- Interinstrument differences were observed and were more attributable to individual instruments than manufacturers.
- Survey specimens, being buffered aqueous solutions, may not fully represent blood characteristics, potentially creating artifacts, especially for PO2.
Conclusions:
- The study highlights variability in blood gas analysis, particularly for PO2, among clinical laboratories.
- Instrument-specific factors contribute more to measurement differences than manufacturer.
- Caution is advised when interpreting survey results due to the limitations of the aqueous survey material.
- Further development of reference methods is needed for accurate assessment of blood gas analysis performance.