On-column propylation method for measuring plasma valproate concentration by gas chromatography
Therapeutic Drug Monitoring
|January 1, 1981
Summary
This study presents a fast and accurate gas-chromatographic method for measuring plasma valproate concentrations. The technique uses double-solvent extraction and on-column propylation, making it ideal for routine laboratory analysis.
Area of Science:
- Analytical Chemistry
- Clinical Chemistry
- Pharmacokinetics
Background:
- Accurate measurement of plasma valproate is crucial for therapeutic drug monitoring.
- Existing methods may be time-consuming or complex for routine use.
Purpose of the Study:
- To develop and validate a rapid, accurate, and simple gas-chromatographic method for quantifying plasma valproate concentrations.
Main Methods:
- Gas-chromatography with on-column propylation.
- Double-solvent extraction of plasma samples.
- Use of 2-propylhexanoic acid as an internal standard.
- Confirmation of derivative identity using gas chromatography-mass spectrometry.
Main Results:
- The method demonstrated accuracy and simplicity.
- Duplicate plasma sample analysis was achievable within 30 minutes.
- Gas chromatography-mass spectrometry confirmed the identity of propyl ester derivatives.
Conclusions:
- The developed gas-chromatographic method is suitable for routine laboratory use.
- The assay's speed and accuracy facilitate efficient therapeutic drug monitoring of valproate.