Semi-automatic blood lactate assay: experiences from an exercise laboratory
International Journal of Sports Medicine
|February 1, 1983
Summary
This study shows that a semi-automatic flow injection analysis (FIA) system accurately measures blood lactate concentrations in exercising subjects. Non-neutralized samples are reliable, enabling faster testing in exercise laboratories.
Area of Science:
- Biochemistry
- Sports Science
- Analytical Chemistry
Background:
- Blood lactate is a key indicator in exercise physiology.
- Accurate and rapid lactate measurement is crucial for exercise monitoring.
- Existing manual methods can be time-consuming.
Purpose of the Study:
- To evaluate a semi-automatic flow injection analysis (FIA) system for blood lactate determination.
- To compare lactate measurements from non-neutralized and neutralized blood samples.
- To validate FIA lactate results against established manual enzymatic and colorimetric methods.
Main Methods:
- Utilized a semi-automatic flow injection analysis (FIA) system coupled with fluorometric enzymatic methods.
- Collected duplicate blood samples from exercising subjects.
- Compared FIA results with manual enzymatic and colorimetric assays.
- Assessed lactate concentrations in both neutralized and non-neutralized samples.
Main Results:
- The FIA method demonstrated high reliability with a coefficient of variation of 4.9%.
- No significant difference in lactate concentration was found between non-neutralized and neutralized samples.
- FIA lactate values were validated against two manual assay methods.
- The FIA system achieved a rapid throughput of 60 samples per hour.
Conclusions:
- The semi-automatic FIA system provides a rapid, accurate, and reliable method for blood lactate determination.
- Non-neutralization of samples is a viable option, enhancing sample turnover efficiency.
- This FIA method is highly relevant for exercise laboratories requiring fast and precise lactate analysis.