Crystal-controlled telemetry system for long-term acquisition of physiological data.
The American Journal of Physiology
|January 1, 1983
Summary
A new crystal-controlled telemetry system enables reliable, long-term blood pressure monitoring in unrestrained animals. This high-dependability system offers continuous, unattended data recording for over 200 hours.
Area of Science:
- Biomedical Engineering
- Physiological Monitoring
Background:
- Accurate, long-term blood pressure monitoring in unrestrained animals is crucial for research.
- Existing methods may be invasive or limited in duration and reliability.
Purpose of the Study:
- To develop a highly reliable, crystal-controlled telemetry system for continuous blood pressure recording.
- To enable unattended, around-the-clock data acquisition in animal models.
Main Methods:
- Development of a miniaturized transducer-transmitter system powered by lithium batteries.
- Utilized a novel piezoresistive blood pressure transducer for enhanced dependability.
- Employed crystal-controlled radio-frequency oscillators for stable signal transmission and reception.
Main Results:
- The system provides over 200 hours of continuous blood pressure recording.
- Achieved a flat frequency response (0-25 Hz) with time stability of +/- 1 mmHg/day.
- Demonstrated linearity better than 0.5% and highly reliable signal reception through frequency locking.
Conclusions:
- The developed crystal-controlled telemetry system offers high dependability for long-term blood pressure monitoring.
- Facilitates continuous, unattended physiological data acquisition in unrestrained animals.
- Presents a significant advancement for cardiovascular research in animal models.