Effects of sterilization on periodontal instruments
Journal of Periodontology
|July 1, 1982
Summary
Sterilization methods like steam, formalin-alcohol vapor, and dry heat do not visually damage stainless steel periodontal instruments. However, saturated steam can cause oxidation and dulling of carbon steel instruments.
Area of Science:
- Dental Instruments
- Materials Science
- Sterilization Technologies
Background:
- The sharpness of periodontal instruments is critical for effective treatment.
- Understanding the impact of sterilization on instrument integrity is essential for clinical practice.
Purpose of the Study:
- To evaluate the effects of common sterilization methods on the cutting edges of periodontal curets and scalers.
- To compare the durability of stainless steel versus carbon steel instruments after repeated sterilization cycles.
Main Methods:
- New, sharpened stainless steel and carbon steel instruments were subjected to 10 sterilization cycles.
- Sterilization methods included saturated steam, formalin-alcohol vapor, and dry heat at specified temperatures.
- Instruments were examined using scanning electron microscopy (SEM) before and after sterilization.
Main Results:
- No visual damage was observed on stainless steel instruments sterilized by saturated steam (250°F), formalin-alcohol vapor (270°F), or dry heat (up to 340°F).
- Carbon steel instruments showed no damage after formalin-alcohol vapor (270°F) or dry heat (up to 340°F) sterilization.
- Saturated steam sterilization (250°F) resulted in oxidation and dulling of the cutting edges of carbon steel instruments.
Conclusions:
- Stainless steel periodontal instruments demonstrate excellent resistance to damage from common sterilization methods.
- Carbon steel instruments are susceptible to damage from saturated steam sterilization, impacting their cutting edge integrity.
- Careful selection of sterilization methods is crucial for maintaining the performance of carbon steel periodontal instruments.