Evaluation of the pneuma-tonometer

S Wittenberg

    American Journal of Optometry and Physiological Optics
    |May 1, 1978
    PubMed
    Summary

    This study compared Goldmann applanation tonometry and Pneuma-Tonometer for measuring intraocular pressure. The Pneuma-Tonometer showed good intra-reader reliability but requires calibration adjustments for accurate intraocular pressure assessment.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Characteristics of patients with epithelioid hemangioendothelioma (EHE): a retrospective analysis of the Charité- Universitätsmedizin Berlin.

    Journal of cancer research and clinical oncology·2026
    Same author

    Digital adaptation of teaching disaster and deployment medicine under COVID-19 conditions: a comparative evaluation over 5 years.

    BMC medical education·2022
    Same author

    [Periprosthetic fractures of the acetabulum: osteosynthesis].

    Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen·2020
    Same author

    Acetabular Fractures in an Advanced Age - Current Knowledge and Treatment Options.

    Acta chirurgiae orthopaedicae et traumatologiae Cechoslovaca·2017
    Same author

    Fixation of acetabular fractures via the ilioinguinal versus pararectus approach: a direct comparison.

    The bone & joint journal·2015
    Same author

    Comment on: Laser Speckle for Determining Ametropia and Accommodative Response of the Eye.

    Applied optics·2010

    Area of Science:

    • Ophthalmology
    • Biomedical Engineering

    Background:

    • Intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
    • Goldmann applanation tonometry is the gold standard, but alternative devices are sought for efficiency and accessibility.

    Purpose of the Study:

    • To evaluate the reliability and agreement of the Pneuma-Tonometer against the Goldmann applanation tonometer for IOP measurement.
    • To assess the intra-reader variability of the Pneuma-Tonometer.

    Main Methods:

    • 125 patients undergoing routine eye care had their IOP measured using both Goldmann applanation tonometer and Pneuma-Tonometer.
    • Pneuma-Tonometer readings were independently assessed by two blinded observers.
    • Statistical analysis included standard deviation of differences and correlation coefficients.

    Main Results:

    • The Pneuma-Tonometer demonstrated low intra-reader variability (SD=0.44 mm Hg).
    • Paired Pneuma-Tonometer readings showed higher variability (SD=1.75 mm Hg).
    • The Pneuma-Tonometer readings were, on average, 0.40 mm Hg higher than Goldmann readings, with a strong correlation (r=0.88) and a standard deviation of differences of 1.50 mm Hg.

    Conclusions:

    • The Pneuma-Tonometer exhibits good intra-reader reliability.
    • Calibration adjustments are necessary for the Pneuma-Tonometer to align with Goldmann applanation tonometer measurements.
    • Alternative calibration methods beyond linear regression may be beneficial.