Related Experiment Videos
Comparative evaluation of coated and uncoated polyglactin 910 in cataract and muscle surgery
Abstract:
This study describes the intraoperative and postoperative performance of coated and uncoated Polyglactin 910 in cataract and muscle surgery. The evaluation represents an animal and clinical comparison of 5-0 coated and uncoated Polyglactin 910 in 50 muscle cases and 7-0 coated and uncoated Polyglactin 910 in 150 cataract cases. Both Polyglactin 910 sutures provided high tensile strength, good knot security, batch-to-batch uniformity, ease of handling, minimal reaction, and a predictable absorption rate. Improvements noted in the coated suture were: (1) easier passage through tissue, (2) less chatter in tying, (3) fewer instances of premature tie, and (4) decreased tendency to incarcerate tissue.