Related Experiment Video
Updated: Aug 15, 2026

Construction and Implantation of a Microinfusion System for Sustained Delivery of Neuroactive Agents.
Published on: March 18, 2008
A portable insulin infusion system with a rotary solenoid-driven peristaltic pump
Abstract:
A portable, external, preprogrammed insulin infusion system with a mass of 560 grams and a battery lifetime of 4 months is described. The system uses a rotary solenoid-driven peristaltic pump, with low-power CMOS timing circuitry and controls to provide bimodal insulin delivery with a timed high rate. Accurate and reproducible delivery of 2 microliter fluid pulses at programmed intervals from 0.6 seconds to 100 minutes can be achieved permitting the delivery of a wide range of insulin concentrations. This system is the precursor to a totally implantable insulin delivery system.
Related Concept Videos
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into rapid-acting...