Electroretinography with noncorneal and corneal electrodes
Investigative Ophthalmology & Visual Science
|November 1, 1978
Summary
Noncorneal electroretinography (ERG) provides an alternative when corneal recording is not possible. An electrode at the medial canthus with medial viewing closely correlates with standard corneal ERG measurements.
Area of Science:
- Ophthalmology
- Neuroscience
- Medical instrumentation
Background:
- Corneal electroretinography (ERG) is a standard diagnostic tool.
- Certain patient conditions contraindicate corneal electrode placement.
- Noncorneal ERG methods are needed as alternatives.
Purpose of the Study:
- To investigate noncorneal electroretinography (ERG) montages.
- To assess the correlation between noncorneal and corneal ERG recordings.
- To identify the optimal noncorneal montage for clinical use.
Main Methods:
- Three noncorneal electroretinography (ERG) montages were evaluated.
- Correlation analysis was performed between noncorneal and corneal ERG signals.
- Repeatability of each noncorneal montage was assessed.
Main Results:
- An electrode at the medial canthus combined with medial viewing showed the highest correlation with corneal ERG.
- All tested noncorneal montages demonstrated good repeatability.
- The inner canthus electrode montage was less susceptible to time-locked artifacts.
Conclusions:
- Noncorneal electroretinography (ERG) using an inner canthus electrode is a viable alternative to corneal ERG.
- This method correlates well with corneal ERG and can be used with electro-oculography.
- Medial canthus electrode placement offers a reliable option for patients unable to undergo corneal recording.