An interlaboratory comparison of serum total protein analyses
American Journal of Clinical Pathology
|September 1, 1978
Summary
The 1976 Comprehensive Chemistry Survey found biuret reaction most common for total serum protein measurement. While precisions were comparable across labs, the SMAC system showed a negative bias in concentration results.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Analytical Chemistry
Background:
- The College of American Pathologists (CAP) Comprehensive Chemistry Survey is a vital tool for laboratory quality assessment.
- Accurate measurement of total serum proteins is crucial for diagnosing various medical conditions.
Purpose of the Study:
- To analyze the data from the 1976 CAP Comprehensive Chemistry Survey regarding total serum protein measurements.
- To evaluate the precision and accuracy of different analytical systems and methods used for total serum protein estimation.
Main Methods:
- Analysis of data contributed by over 2,000 laboratories participating in the 1976 CAP survey.
- Evaluation of total serum protein concentration using the biuret reaction and refractometry as primary methods.
- Assessment of precision and accuracy data across various analytic systems.
Main Results:
- The biuret reaction was the most prevalent method for total serum protein estimation, followed by refractometry.
- Precision among participating laboratories was found to be generally comparable across different analytical systems.
- A notable finding was a negative bias in total protein concentration measurements reported by the SMAC system.
Conclusions:
- The biuret method remains a widely adopted and reliable technique for total serum protein analysis.
- While overall precision is good, specific analytical systems like SMAC may introduce biases that require attention.
- The survey highlights the importance of continuous monitoring and evaluation of laboratory methods for accurate patient diagnosis.