A new computerized method for hearing aid fitting based on measurements at the ear drum
Scandinavian Audiology
|January 1, 1981
Summary
A new computerized system measures hearing aid etymotic frequency characteristics. Standard measurements differ significantly from ear canal readings, making optimization challenging.
Area of Science:
- Audiology
- Medical Instrumentation
- Acoustics
Background:
- Hearing aid fitting relies on standardized coupler measurements (IEC 118, 2 cc).
- These measurements may not accurately reflect the actual sound delivered to the eardrum (etymotic).
- Accurate etymotic measurements are crucial for effective hearing aid performance.
Purpose of the Study:
- To introduce a computerized system for measuring hearing aid etymotic frequency characteristics.
- To compare etymotic measurements with standard coupler measurements.
- To assess the feasibility of optimizing hearing aid fittings based on etymotic characteristics.
Main Methods:
- Development of a computerized system for in-situ etymotic frequency characteristic measurement.
- Utilizing a probe microphone placed at the eardrum.
- Comparison with measurements using a standard 2 cc coupler (IEC 118).
- Clinical investigation involving patients and various hearing aids/earmolds.
Main Results:
- Significant discrepancies observed between standard coupler measurements and etymotic measurements.
- Difficulty in practically optimizing the etymotic frequency characteristic by adjusting the hearing aid or earmold.
- Limited variation in the upper limiting frequency when using different hearing aids with the same earmold on the same subject.
Conclusions:
- Standard hearing aid coupler measurements do not reliably predict the etymotic frequency response.
- Current methods for optimizing hearing aid fittings may be inadequate.
- Further research is needed to refine in-situ measurement techniques and fitting strategies for improved audiological outcomes.