A candidate reference method for determination of total protein in serum. II. Test for transferability
Clinical Chemistry
|October 1, 1981
Summary
This study validated a total serum protein measurement method across multiple labs. The reference method demonstrated good transferability and precision for accurate protein quantification.
Area of Science:
- Clinical Chemistry
- Biotechnology
- Analytical Chemistry
Background:
- Accurate measurement of total serum protein is crucial for clinical diagnostics.
- Standardization of analytical methods is essential for reliable laboratory results.
- The candidate Reference Method for total serum protein required validation for inter-laboratory transferability.
Purpose of the Study:
- To assess the transferability of a candidate Reference Method for total serum protein.
- To evaluate the method's performance in multiple independent laboratories.
- To establish the reliability of the method for routine clinical use.
Main Methods:
- The study involved eight laboratories in the United States and Europe.
- National Bureau of Standards SRM 927 (bovine serum albumin) was used as the calibration standard.
- Four serum pools were analyzed over 15 days to assess variation.
Main Results:
- The mean absorptivity value for the calibration standard was 0.2983 L g-1 cm-1.
- Within-run coefficients of variation (CV) ranged from 0.1% to 2.5%.
- Day-to-day (total) variation was between 0.4% and 1% in most laboratories.
Conclusions:
- The candidate Reference Method for total serum protein is transferable across different laboratories.
- The method exhibits acceptable precision for reliable protein quantification.
- This validation supports the adoption of the method for standardized clinical laboratory testing.