Evaluation of home glucose measuring devices
The Medical Journal of Australia
|August 22, 1981
Summary
Home glucose monitoring systems vary in accuracy. While some strips perform well with experienced users, machine-based systems reduce errors. Training and regular checks are crucial, especially for patients with high blood glucose levels.
Area of Science:
- Clinical Chemistry
- Medical Devices
- Diabetes Management
Background:
- Numerous home glucose-monitoring systems are available for diabetes self-management.
- Evaluating the accuracy and reliability of these devices is crucial for patient care.
Purpose of the Study:
- To assess the accuracy, precision, and variability of various home glucose-monitoring systems.
- To determine the suitability of different systems for patient use in home glucose monitoring.
Main Methods:
- Comparison of whole blood glucose measurements from home systems against a reference method in a clinical setting.
- Evaluation of accuracy (mean difference) and precision (standard deviation of differences).
- Assessment of model and operator variability for each system.
Main Results:
- The 20-800 BM Glycemia strips showed good performance with experienced operators but high interoperator variability.
- Machine-based systems demonstrated reduced interoperator differences compared to strip-only methods.
- The Glucochek Mk I system did not meet satisfactory performance standards.
- All tested systems exhibited decreased accuracy and precision at blood glucose levels exceeding 15.0 mmol/L.
Conclusions:
- A dedicated machine is not essential for all home glucose-monitoring systems.
- Patient training and regular verification of results against a reference method are vital for reliable home monitoring.
- Home glucose monitoring is unreliable in patients with significant hyperglycemia.