A comparison study of blood glucose test strips
Diabetes Care
|May 1, 1981
Summary
This study compared blood glucose test strips, finding BM Glycemie 20-800 (Chemstrip bG) more accurate visually. For meter systems, Reflotest-Glucose offered superior accuracy and precision over Dextrostix.
Area of Science:
- Clinical Chemistry
- Medical Diagnostics
Background:
- Blood glucose test strips are essential for rapid glucose estimation.
- Visual and meter-based readings are common methods for interpreting test strips.
Purpose of the Study:
- To compare the performance characteristics of principal blood glucose test strips.
- To evaluate accuracy and precision of visual and meter-based readings.
Main Methods:
- Comparative performance analysis of available blood glucose test strips.
- Visual assessment against color scales.
- Evaluation using reflectance meter systems.
Main Results:
- BM Glycemie 20-800 (Chemstrip bG) demonstrated higher accuracy in visual readings compared to Dextrostix.
- Reflotest-Glucose showed greater accuracy and precision than Dextrostix when used with a meter.
- Reflotest systems exhibited stable color development for at least two days, unlike Dextrostix.
- Reflotest required two strips to cover the full blood glucose range with a meter.
Conclusions:
- Visual interpretation favors BM Glycemie 20-800 (Chemstrip bG) for accuracy.
- Reflotest-Glucose offers enhanced accuracy and precision for meter-based blood glucose monitoring.
- Color stability is a key advantage of Reflotest systems.