Limbal pneumatonometry

J M Breitfeller, D L Krohn

    American Journal of Ophthalmology
    |March 1, 1980
    PubMed
    Summary

    This study compared Goldmann applanation tonometry and Langham pneumatonometry for measuring intraocular pressure. Pneumatonometry readings were higher, but conversion factors were developed for predicting Goldmann values, aiding diagnosis with corneal disease.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Repair of bilateral common iliac artery aneurysms coexisting with a pelvic horseshoe kidney.

    Annals of vascular surgery·1999
    Same author

    Emergency! Peritonitis.

    The American journal of nursing·1999
    Same author

    The effect of nonporous PTFE-covered stents on intimal hyperplasia following balloon arterial injury in minipigs.

    Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery·1998
    Same author

    Degradation of the Ferric Chelate of EDTA by a Pure Culture of an Agrobacterium sp.

    Applied and environmental microbiology·1990
    Same author

    Lectin-mediated attachment of liposomes to cornea: influence on transcorneal drug flux.

    Investigative ophthalmology & visual science·1982
    Same author

    Liposomes in topical drug delivery.

    Investigative ophthalmology & visual science·1982

    Area of Science:

    • Ophthalmology
    • Biomedical Engineering

    Background:

    • Intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
    • Goldmann applanation tonometry is the gold standard, but can be affected by corneal properties.
    • Pneumatonometry offers an alternative IOP measurement method.

    Purpose of the Study:

    • To compare Goldmann applanation tonometry with Langham pneumatonometry in normal and glaucomatous eyes.
    • To develop conversion values for correlating pneumatonometry and applanation tonometry measurements.
    • To assess the potential of pneumatonometry for IOP determination in the presence of corneal disease.

    Main Methods:

    • A comparative study involving Goldmann applanation tonometry and Langham pneumatonometry (central and limbal) was conducted.
    • Measurements were performed on normal and glaucomatous eyes with normal corneas, in both sitting and supine positions.
    • Standardized measurement sequences were applied during a single visit.

    Main Results:

    • Mean pneumatonograph measurements were significantly higher than corresponding Goldmann applanation values.
    • No significant difference was observed between limbal and apical pneumatonograph values in the supine position.
    • Clinical Correlation Estimators were derived to predict Goldmann applanation values from limbal pneumatonometry.

    Conclusions:

    • Pneumatonometry provides higher IOP readings than Goldmann applanation tonometry.
    • Derived conversion values allow for prediction of Goldmann applanation tonometry values with 95% confidence.
    • This method shows promise for IOP assessment in patients with corneal abnormalities.