Ipsilateral acoustic reflex artefacts measured in cadavers
Scandinavian Audiology
|January 1, 1980
Summary
This study found that most impedance meters produced artifactual acoustic reflex indications in cadaver ears. A pulsed stimulus technique showed fewer artifacts, suggesting improvements for audiological testing.
Area of Science:
- Audiology
- Biomedical Engineering
- Otoacoustic Emissions
Background:
- Impedance audiometry is crucial for assessing middle ear function.
- Acoustic reflexes are commonly measured using impedance meters.
- Artifacts can compromise the accuracy of impedance measurements.
Purpose of the Study:
- To determine the minimum stimulus levels causing artifactual impedance changes in cadaver ears.
- To evaluate the effectiveness of different impedance meter designs in minimizing artifacts.
- To investigate methods for increasing artifact thresholds.
Main Methods:
- Six impedance meters were used to present acoustic stimuli to cadaver ears.
- Stimulus levels and frequencies were varied to detect impedance changes.
- Pressure differentials across the tympanic membrane were manipulated.
Main Results:
- Artifactual impedance changes were detected in five of six instruments across tested frequencies.
- One instrument using a pulsed stimulus technique exhibited significantly fewer artifacts.
- Introducing pressure differentials yielded only minor increases in artifact thresholds for most devices.
Conclusions:
- Most conventional impedance meters are susceptible to generating artifacts during acoustic reflex testing.
- Pulsed stimulus techniques show promise in reducing measurement artifacts.
- Further research is needed to optimize impedance measurement techniques and minimize artifacts in audiological assessments.