Anaesthetic machine pipeline inlet pressure gauges do not always measure pipeline pressure
Abstract:
Some anaesthetic gas machines have pipeline inlet pressure gauges which indicate the higher of either pipeline pressure, or machine circuit pressure (the pressure distal to the pressure reducing valve, and proximal to the flowmeter control valve). Failure by the operator to appreciate this feature may in specific circumstances lead to a delayed appreciation of pipeline malfunction or disconnection. The Canadian Standards Association Z168.3-M1980 Anaesthetic Gas Machine Standard requires pipeline inlet gauges which measure only pipeline (hose) pressure. Existing machines should be modified to accommodate this requirement.
Related Concept Videos
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pressure Gauges
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Pipe Flowrate Measurement
The orifice meter is a simple,...
Pipe Flowrate Measurement: Problem Solving