Reproducibility of echocardiography. A study evaluating the variability of serial echocardiographic measurements
Abstract:
In this study, serial reproducibility was determined on standard M-mode (M-mode) and two dimensional guided M-mode (2-D M-mode) records for ten normal volunteers and 20 patients. Overall intraobserver variability in the normal group was 518 +/- 1.8 percent (M-mode) and 3.1 +/- 0.8 percent (2-D M-mode), and in the patient group 3.1 +/- 0.6 percent (M-mode) and 4.7 +/- 0.7 percent (2-D M-mode) (P = NS). Variability on serial examination, however, was twofold to threefold greater. In the normal group, reproducibility varied by 10.8 +/- 1.6 percent (M-mode) and 9.4 +/- 1.2 percent (2-D M-mode), while in the patient group, it was 8.7 +/- 0.9 percent (M-mode) and 9.4 +/- 0.7 percent (2-D M-mode). The lowest serial variability achieved was for the diastolic LV dimension on serial M-mode examination (4 percent); the largest variability pertained to measurement of the interventricular septum (14 percent). Other structures had intermediate variation in reproducibility. This study emphasizes that establishing performance variability for an echocardiographic laboratory is an important part of the interpretation of serial records.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...