[The long-term results of isolated aortic valve replacement with 19 mm prostheses]
T Nakamura1, M Shimamoto, F Yamazaki
1Division of Cardiovascular Surgery, Shizuoka City Hospital, Japan.
Abstract:
Twenty-five patients underwent isolated aortic valve replacement with 19 mm valves (8 cases with St. Jude Medical prostheses, 7 cases with Carpentier Edwards Pericardial prostheses and 10 cases with Björk-Shiley prostheses). Postoperative echocardiography revealed data as follows. Peak pressure gradients (mmHg) were 40.4 +/- 14.4 (mean +/- S.D.) in the SJM group, 23.4 +/- 9.8 in the CEP group and 50.5 +/- 16.7 in the BS group. End-diastolic left ventricular dimension in the BS group and interventricular septum thickness in the CEP group was reduced significantly. All but one case were in NYHA functional class I or II, and clinical improvement was almost satisfactory in all groups. However, in the BS group, 4 cases showed peak pressure gradients more than 50 mmHg. We conclude that 19 mm aortic prostheses in this study can provide satisfactory results. However, as for mechanical valve, it may be undesirable to choose BS prosthesis.