Albuterol delivery in a model of mechanical ventilation. Comparison of metered-dose inhaler and nebulizer efficiency
Related Experiment Video
Updated: Aug 14, 2026

Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
P Diot1, L Morra, G C Smaldone
1Department of Medicine, State University of New York at Stony Brook 11794-8172, USA.
Abstract:
Using an in vitro model, we compared efficiencies of jet nebulizers and metered-dose inhalers (MDI) with actuator devices to deliver albuterol in various conditions of mechanical ventilation. Factors tested included influence of humidification, MDI actuator device (Aerovent spacer or Marquest 172275 MDI adaptor), and synchronization of MDI to the respiratory cycle. With the nebulizer (AeroTech II) filled with 2.5 mg albuterol sulfate in 3 ml water and run until dry, inhaled mass was 42 +/- 2.6% and mass median aerodynamic diameter (MMAD) was 1.3 microns on a nonhumidified circuit. With the MDI+ Aerovent, the inhaled mass percentage per 90 microgm puff was 15.4 +/- 0.2% with humidification and 25.1 +/- 3.7% without humidification, actuations being synchronized with the beginning of inspiration and separated by a 1-min pause. Failure to synchronize actuations with inspiration or to pause 1 min resulted in significant reductions in inhaled mass (by 35 and 72%, respectively). The Marquest adaptor was less efficient, with an inhaled mass of 7.2 +/- 0.7% under optimal conditions. Metered-dose inhaler actuation during expiration resulted in relatively large particles (MMAD = 2.0 microns). All other MDI actuations led to essentially biphasic distributions, with particles greater than 1 micron following a distribution similar to the nebulizer and the overall MMAD estimated to be 0.22 microns. The AeroTech II delivered a cumulative 1,000 micrograms of drug (2,500 x 0.40) over 40 min. To achieve that amount, the MDI connected to the Aerovent and used in its most efficient sequence would require 45 timed puffs (90 micrograms per puff, 25.1% mean inhaled mass) and take 45 min of an experienced therapist's time.(ABSTRACT TRUNCATED AT 250 WORDS)
Related Concept Videos
Inhaled Medications
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)