Determination of spiral CT slice sensitivity profiles using a point response phantom

W J Davros1, B R Herts, J J Walmsley

  • 1Cleveland Clinic Foundation, Division of Radiology Hb-6, OH 44195, USA.

Journal of Computer Assisted Tomography
|September 1, 1995
PubMed
Summary

Optimizing spiral CT scanning involves balancing narrower collimation with higher pitch. This combination reduces partial volume averaging and improves slice sensitivity profiles (SSP) without sacrificing z-axis coverage.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Multi-vendor multi-site quantitative MRI analysis of cartilage degeneration 10 Years after anterior cruciate ligament reconstruction: MOON-MRI protocol and preliminary results.

Osteoarthritis and cartilage·2022
Same author

Donor-matched comparison of chondrogenic progenitors resident in human infrapatellar fat pad, synovium, and periosteum - implications for cartilage repair.

Connective tissue research·2019
Same author

Primary Cells Isolated from Human Knee Cartilage Reveal Decreased Prevalence of Progenitor Cells but Comparable Biological Potential During Osteoarthritic Disease Progression.

The Journal of bone and joint surgery. American volume·2018
Same author

Histopathological assessment of primary osteoarthritic knees in large patient cohort reveal the possibility of several potential patterns of osteoarthritis initiation.

Current research in translational medicine·2017
Same author

Meniscus treatment and age associated with narrower radiographic joint space width 2-3 years after ACL reconstruction: data from the MOON onsite cohort.

Osteoarthritis and cartilage·2015
Same author

Comparison of radiation dose and image quality from single-energy and dual-energy CT examinations in the same patients screened for hepatocellular carcinoma.

Clinical radiology·2014

Area of Science:

  • Medical Imaging
  • Radiology
  • Computed Tomography

Background:

  • Slice sensitivity profiles (SSP) are crucial for assessing CT image quality.
  • Partial volume averaging can degrade image resolution in spiral CT.
  • Optimizing scan parameters is essential to minimize artifacts and enhance diagnostic accuracy.

Purpose of the Study:

  • To compare slice sensitivity profiles (SSP) for various spiral CT scan parameter combinations.
  • To identify optimal settings for reducing partial volume averaging.
  • To ensure no compromise in z-axis coverage during CT scans.

Main Methods:

  • A point response phantom with a 0.4 mm copper ball bearing was utilized.
  • Scans were performed across a range of pitches (0.1-2.0) and collimations (5, 8, 10 mm).

Related Experiment Videos

  • Full width half maximum (FWHM) and full width tenth maximum (FWTM) were calculated from SSP curves.
  • Main Results:

    • Significant differences in FWHMs and FWTMs were observed between different collimations at constant table speeds (p < 0.0001).
    • A linear trend showed increasing FWHMs and FWTMs with higher pitch at constant collimation (p < 0.0013).
    • Narrower collimation and higher pitch resulted in narrower SSPs.

    Conclusions:

    • Scanning with narrower collimation and higher pitch yields improved SSP.
    • This parameter combination effectively reduces partial volume averaging.
    • Z-axis coverage is maintained without compromise.