The "low intensity treadmill exercise" protocol for appropriate rate adaptive programming of minute ventilation
T Lewalter1, D MacCarter, W Jung
1Department of Cardiology, University of Bonn, Germany.
Pacing and Clinical Electrophysiology : PACE
|July 1, 1995
Summary
Rate adaptive pacemakers use minute ventilation to match heart rate to patient effort. A low-intensity treadmill exercise protocol effectively substitutes for peak exercise tests in programming pacemaker response factors.
Area of Science:
- Cardiology
- Exercise Physiology
Background:
- Rate adaptive pacemakers aim to mimic physiological heart rate responses during exercise.
- Minute ventilation (MV) measured by transthoracic impedance is a key parameter for pacemaker rate control.
- Accurate programming of pacemaker response requires understanding the heart rate to minute ventilation (HR/VE) relationship.
Purpose of the Study:
- To determine the physiological HR/VE relationship during peak exercise.
- To develop a database for programming minute ventilation-controlled pacemakers.
- To evaluate if a low-intensity exercise test (LITE) can substitute for peak exercise tests (RITE) in assessing the HR/VE slope below the anaerobic threshold.
Main Methods:
- 41 healthy subjects underwent treadmill exercise testing (RITE and LITE protocols).
- Breath-by-breath gas exchange was collected.
- Linear regression analysis determined HR/VE slopes below and above the anaerobic threshold and during LITE.
Main Results:
- The HR/VE relationship during RITE was curvilinear, not linear.
- The HR/VE slope was steeper below the anaerobic threshold (1.54 ± 0.51) than above (1.15 ± 0.37).
- The HR/VE slope during LITE (1.58 ± 0.88) did not differ significantly from the RITE slope below the anaerobic threshold (P = 0.79).
Conclusions:
- Rate adaptive pacing should simulate the curvilinear HR/VE relationship from rest to peak exercise.
- The LITE protocol effectively substitutes for peak exercise tests in determining the HR/VE slope below the anaerobic threshold.
- LITE enables accurate pacemaker programming for physiological rate response matching patient effort.