Output factors for irregularly shaped electron fields

R G Kurup1, G P Glasgow, L Leybovich

  • 1Department of Radiation Oncology, University of Kansas Cancer Center, Kansas City 66160-7321, USA.

Medical Dosimetry : Official Journal of the American Association of Medical Dosimetrists
|January 1, 1995
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Effectiveness of compensating filters in the presence of tissue inhomogeneities.

Journal of applied clinical medical physics·2003
Same author

Matching tomographic IMRT fields with static photon fields.

Medical physics·2002
Same author

Elimination of field size dependence of enhanced dynamic wedge factors.

Physics in medicine and biology·2000
Same author

Effect of metal reconstruction plates on cobalt-60 dose distribution: a predictive formula and clinical implications.

International journal of radiation oncology, biology, physics·1999
Same author

Technique charts for Kodak's new film-screen systems for portal localization.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·1998
Same author

High dose-rate brachytherapy treatment delivery: report of the AAPM Radiation Therapy Committee Task Group No. 59.

Medical physics·1998
Same journal

Predictive dosimetric parameters for adaptive replanning in proton beam therapy for thoracic cancer.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

A case study of inhale and exhale split-course VMAT treatment to minimize dose to ribs.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

Prospective clinical implementation of the robust optimization planning technique in long-course photon radiotherapy treatment for lung cancer.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

Dosimetric comparison of target coverage, dose gradient, and low-dose bath in left-sided accelerated partial breast irradiation: CyberKnife versus conventional linac-based noncoplanar VMAT and 3D-CRT.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

Cross-prescription validation of knowledge-based planning in prostate VMAT using a structured validation framework.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
Same journal

Dosimetric evaluation of beam block techniques in helical tomotherapy for extranodal nasal-type NK/T-cell lymphoma.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists·2026
See all related articles

Accurate electron beam dosimetry requires output factors for irregular fields. This study presents a practical method using equivalent square fields and interpolation, achieving 1-2% accuracy for electron therapy.

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Radiotherapy Physics

Background:

  • Accurate dose delivery in radiation therapy relies on precise knowledge of output factors for electron beams.
  • Measuring output factors for each patient's irregularly shaped electron fields is time-consuming and impractical.
  • Existing methods for irregular fields may lack efficiency or require complex calculations.

Purpose of the Study:

  • To develop a simple and practical method for determining output factors of irregularly shaped electron beam fields.
  • To reduce the inconvenience of measuring individual output factors for each patient.
  • To provide an accurate estimation of output factors for clinical application in electron beam therapy.

Main Methods:

  • Calculating the equivalent square field area for irregularly shaped electron fields.

Related Experiment Videos

  • Using measured output factors from two standard square fields.
  • Interpolating between known output factors based on equivalent square areas.
  • Main Results:

    • An empirical method was successfully developed to estimate output factors for irregular electron fields.
    • The method utilizes interpolation between output factors of two square fields with known areas.
    • The accuracy of this empirical method was found to be approximately 1% to 2%.

    Conclusions:

    • This interpolation method offers a practical and efficient solution for obtaining output factors for irregularly shaped electron fields.
    • The technique simplifies the process of electron beam dosimetry, enhancing clinical workflow.
    • The achieved accuracy supports its use in routine radiotherapy practice for electron beam treatments.