Related Experiment Videos
Comparison of 26-gauge Atraucan and 25-gauge Whitacre needles: insertion characteristics and complications
S K Sharma1, D R Gambling, G P Joshi
1Department of Anesthesiology and Pain Management, University of Texas, Southwestern Medical Center at Dallas 75235-9068, USA.
Abstract:
Ninety-six women undergoing post-partum tubal ligation under spinal anaesthesia were studied to compare 26G Atraucan with 25G Whitacre spinal needles for ease of insertion, number of attempts at needle insertion, cerebrospinal fluid (CSF) flow characteristics through the needles, quality of subsequent analgesia, and incidence of perioperative complications. A higher rate of successful dural puncture at the first attempt (40/50 vs 27/46, P < 0.05) and faster (mean +/- SD, 11.5 +/- 2.2 vs 13.5 +/- 2.4, P < 0.001) CSF flow through the needle was achieved with the Atraucan than with the Whitacre needle. The incidence of failed spinal (4% vs 5%) and post-dural puncture headache (PDPH) (4% vs 4.3%) was similar with both needles, but more patients experienced paraesthesiae during needle insertion with the Whitacre than with the Atraucan needle (15% vs 2%, P < 0.05). We conclude that the use of the 26G Atraucan needle is associated with a higher rate of successful identification of the subarachnoid space at the first attempt, faster CSF backflow, and fewer paraesthesia when compared with the 25G Whitacre needle.