Comparison of twenty three nebulizer/compressor combinations for domiciliary use
Related Experiment Video
Updated: Aug 10, 2026

Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
E C Smith1, J Denyer, A H Kendrick
1Respiratory Dept, Bristol Royal Infirmary, UK.
Abstract:
We have assessed the physical and dynamic characteristics of 23 home jet nebulizer/compressor combinations currently available in the UK and Europe. The combinations were evaluated in terms of pressure-flow characteristics, aerosol mass distribution, volume output, electrical costs, and sound level. In addition, we determined the effect of nebulizer fill volume on aerosol mass distribution and volume output. One nebulizer was used with six different compressors, and four compressors were tested with three different nebulizers. The pressure-flow relationships showed a wide variation between models, as did flow-rate at the nebulizer (range 3.0-8.0 L.min-1). The mean +/- SD volume nebulized after 10 min using an initial fill volume of 2.5 and 5.0 mL was 46 +/- 9 and 34 +/- 12%, respectively. The mass median aerodynamic diameter (MMAD) over a 5 min nebulization ranged 2.6 to 10.2 microns. Nine of the nebulizations produced an MMAD of less than 5 microns at both fill volumes. Changing nebulizer/compressor combinations affected flow rate, MMAD and volume output. Sound levels varied between models. Running costs were low, with all using less than 74 kilowatt hours of energy per year. We conclude that there is a wide variation in performance of nebulizer/compressor combinations for use with nebulized bronchodilators. Correct matching of the nebulizer/compressor is seen to be important to ensure optimum performance.(ABSTRACT TRUNCATED AT 250 WORDS)
Related Concept Videos
Inhaled Medications
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)