The twist technique: a method to minimize wire prolapse during Bird's Nest filter placement
1Section of Cardiovascular and Interventional Radiology, Methodist Hospital, Baylor College of Medicine, Houston, TX 77030, USA.
Summary
Applying two or three 360-degree twists during Bird's Nest filter insertion significantly reduces fine-wire prolapse. This technique improves safety and efficacy in inferior vena cava filter placement.
Area of Science:
- Interventional Radiology
- Medical Device Technology
- Vascular Surgery
Background:
- Inferior vena cava (IVC) filters are crucial for preventing pulmonary embolism.
- Bird's Nest filters are widely used but can be associated with complications.
- Fine-wire prolapse, a potential complication, can lead to filter malfunction or vena cava damage.
Purpose of the Study:
- To evaluate the efficacy of a twist technique in reducing fine-wire prolapse.
- To assess the impact of different numbers of twists on prolapse occurrence and severity.
- To optimize Bird's Nest filter insertion procedures.
Main Methods:
- Retrospective analysis of abdominal radiographs following Bird's Nest filter placement.
- Comparison of filter placement with no twists, two 360-degree twists, and three 360-degree twists.
- Quantitative assessment of prolapse length and frequency.
Main Results:
- No twists: 32% prolapse rate (average length 2.8 cm).
- Two twists: 9.1% prolapse rate (P < .005), reduced filter nest length.
- Three twists: 4.9% prolapse rate (P < .005), similar prolapse frequency to two twists.
- No statistically significant difference in prolapse frequency between two and three twists.
Conclusions:
- The twist technique significantly reduces Bird's Nest filter wire prolapse.
- Both two and three 360-degree twists are effective in mitigating this complication.