Related Experiment Videos
3D coronary MR angiography in multiple breath-holds using a respiratory feedback monitor
Y Wang1, R C Grimm, P J Rossman
1Mayo Clinic, Rochester, MN 55905, USA.
Magnetic Resonance in Medicine
|July 1, 1995
Abstract:
To reduce respiratory blur and ghosts in 3D coronary imaging, a data acquisition scheme using consistent multiple breath-holds was implemented. A navigator echo was acquired and processed in real time to dynamically measure diaphragm position. This information was provided as a visual prompt to the patient to maintain consistency in breath-hold levels such that the variation range of diastolic heart position was less than 2 mm. Preliminary results indicate that this multiple breath-hold acquisition scheme, compared with acquisition under respiration, can significantly reduce blur and ghost artifacts in 3D coronary imaging.