Effects of internal defibrillation on an implanted pacing system with programmable polarity
1Department of Medicine, Baylor College of Medicine, Houston, Texas 77030.
Pacing and Clinical Electrophysiology : PACE
|January 1, 1995
Summary
This study found that a specific programmable polarity pacemaker (RelayR 294-03) maintained normal function after defibrillation shocks from an implantable cardioverter defibrillator (ICD) in dogs. Bipolar pacing configurations significantly reduced electrical potentials, indicating safety for combined use.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Managing cardiac arrhythmias often involves implantable cardioverter defibrillators (ICDs) and pacemakers.
- Ensuring pacemaker function after defibrillation shock is critical for patient safety.
Purpose of the Study:
- To evaluate the safety and compatibility of a programmable polarity pacemaker (RelayR 294-03) with an ICD system.
- To assess pacemaker performance following high-energy defibrillation shocks in a canine model.
Main Methods:
- Three RelayR 294-03 pacemakers were tested in dogs with leads in the right atrium and ventricle.
- Pacemakers were subjected to at least 24 high-energy biphasic and monophasic shocks from an ICD system.
- Intracardiac electrical potentials were measured during defibrillation using bipolar and unipolar configurations.
Main Results:
- No pacemaker system failures or mode switching (unipolar pacing/backup mode) were observed post-shock.
- Bipolar pacing configurations (tip-ring) resulted in significantly lower intracardiac potentials compared to unipolar configurations.
- Pacemakers consistently functioned within preshock specifications.
Conclusions:
- The programmable polarity pacemaker (RelayR 294-03) appears safe for use with ICD systems.
- Bipolar pacing configurations mitigate high electrical potentials during defibrillation.
- Further clinical testing is recommended before widespread adoption of combined ICD and pacemaker implantation.