Adaptation of the troponin T ELISA test to a microplate immunoassay reader
I Penttilä1, K Hirvonen, A Julkunen
1Department of Clinical Chemistry, Kuopio University Hospital, Finland.
Abstract:
Troponin T concentration in serum is usually measured by the automated method developed by Boehringer Mannheim for the ES-series of analysers. These instruments need at least 140 microliters of serum and 700 microliters of reagents for a single analysis, which takes 90 min. We describe an alternative procedure, using streptavidin-coated microtitre plates, troponin T reagents of Boehringer Mannheim and an ELISA-reader to measure the concentration of troponin T. The present assay needs only 30 microliters of sample and 200 microliters of reagents, and it takes 75 min; the detection limit is 0.10 microgram/l. We also assessed the microtitre plate method for sensitivity and precision and compared the results with those measured by an ES-300 automatic analyser. Both methods have the same measurement range for troponin T of 0.1 to 15 micrograms/l. For daily routine use of the microtitre plate method we recommend duplicate determinations.