A hemoglobin A1C immunoassay method not affected by carbamylated hemoglobin
Annals of Clinical and Laboratory Science
|January 1, 1995
Summary
Carbamylated hemoglobin (carb Hb) can interfere with Hemoglobin A1C (HbA1C) testing. A new immunoassay method shows no interference from carb Hb, unlike traditional electrophoretic methods, suggesting calibration differences may explain discrepancies.
Area of Science:
- Clinical Chemistry
- Analytical Biochemistry
- Immunodiagnostics
Background:
- Hemoglobin A1C (HbA1C) measurement is crucial for diabetes management.
- Charge-based HbA1C assays are susceptible to interference from carbamylated hemoglobin (carb Hb).
- Carb Hb forms from urea dissociation to cyanate and subsequent reaction with hemoglobin.
Purpose of the Study:
- To evaluate a novel immunoassay for HbA1C, hypothesized to be resistant to carb Hb interference.
- To compare the performance of the immunoassay with an electrophoretic method in the presence of carb Hb.
Main Methods:
- In vitro carbamylation of hemoglobin.
- Electrophoretic separation and densitometry to assess HbA1C and carb Hb co-migration.
- Comparison of HbA1C results from immunoassay and electrophoresis using in vivo samples.
Main Results:
- Electrophoresis showed a 5-7 fold increase in HbA1C peak with carbamylation, while immunoassay results remained unaffected.
- In vivo studies revealed a proportional bias (y = 1.2x - 0.21 percent) favoring the electrophoretic system.
- Electrophoretic methods exhibited significant interference from carb Hb, unlike the tested immunoassay.
Conclusions:
- The novel immunoassay demonstrates robustness against carb Hb interference.
- Observed discrepancies between methods may be attributed to calibration issues rather than solely carb Hb interference.
- This finding has implications for accurate HbA1C monitoring in patients with potential Hb modifications.